SOP for Operational Check of Friability Test Apparatus
1.0 OBJECTIVE
To calibrate the equipment for reliable and accurate results.
2.0 SCOPE
This procedure is applicable to calibrate the friability apparatus, installed in quality control laboratory.
[ads id="ads1"]
3.0 RESPONSIBILITY
3.1 Doing: Technical Assistant
3.2 Checking: Executive /Manager
4.0 ACCOUNTABILITY
Head of the Department
5.0 PROCEDURE
Frequency: Once in a three month
5.1 Operate the instrument as per SOP.
5.2 Set the timer ‘4 min’.
5.2.1 Start the instrument and check the time against a stopwatch.
[ads id="ads1"]
5.2.2 Enter the results in Annexure-I
5.3.1 Start the instrument and note down the number of revolution per minutes.
5.4 Set the revolution at 100.
5.4.1 Start the instrument and note down the number of revolution.
5.5 Record all results in Annexure-I.
5.6 If results are out of tolerance limit, follow the SOP.
[ads id="ads1"]
6.0 ABBREVIATION
6.1 SOP - Standard Operating Procedure