SOP for Calibration of Brook Field Viscometer

1.0 OBJECTIVE

To calibrate the equipment for reliable and accurate results.

2.0 SCOPE

This procedure is applicable to the Brook field viscometer, installed in the quality control department.

[ads id="ads1"]

3.0 RESPONSIBILITY

3.1 Doing: Technical Assistant

3.2 Checking: Executive /Manager

4.0 ACCOUNTABILITY

Head of the Department

[ads id="ads1"]

5.0 PROCEDURE

Frequency of calibration: Once in three month

5.1 Operate the instrument as per SOP

5.2 Calibrate the instrument with standard viscosity solution of 10, 100, 500, 5000 & 12,500 cps using respective spindles and record the results in Annexure-I

5.3 The performance of the instrument is satisfactory, if the measured viscosity is within the given tolerance limit, otherwise follow the S.O.P.

[ads id="ads1"]

6.0 ABBREVIATION

6.1 0C = Centigrade

6.2 S.O.P. = Standard operating procedure

6.3 cps = Centipoise

[ads id="ads1"]