SOP for Calibration of Gas Chromatograph

1.0 OBJECTIVE

To check the performance of equipment for reliable and accurate results.

2.0 SCOPE

This procedure is applicable to check the performance of Gas chromatograph, installed in the quality control department.

[ads id="ads1"]

3.0 RESPONSIBILITY

3.1 Doing: Technical Assistant

3.2 Checking: Executive /Manager

4.0 ACCOUNTABILITY

Head of the Department

5.0 PROCEDURE

Frequency: Once in three months

5.1 Operate the instrument as per SOP

5.2 Prepare the solutions as follows.

[ads id="ads1"]

5.2.1 INTERNAL STANDARD: Dilute 25 ml of Ethanol to 1000 ml with purified water.

5.2.2 TEST PREPARATION: Dilute each of 1ml, 2ml, 3ml, 4ml, 5ml methanol to 100ml with internal standard.

5.3 Follow the GC parameters as per given in Annexure I.

5.4 Inject 2.0 ml of test preparation and record the chromatographs. 

5.5 Calculate the ratio of the area of Methanol v/s area of Internal standard

5.6 Calculate the Correlation Coefficient

5.6.1 Correlation Coefficient should be more than 0.9900.

5.6.2 The performance of the instrument is satisfactory if the obtained Correlation Coefficient is within the given limit, otherwise follow the S.O.P.

[ads id="ads1"]

6.0 ABBREVIAITON

6.1 % = Percentage 

6.2 ml = Mililiter

6.3 v/s = verses 

6.4 GC = Gas Chromatograph

[ads id="ads1"]