SOP for Calibration of High Performance Thin Layer Chromatography

1.0 OBJECTIVE

To check the performance of High performance thin layer chromatography.

2.0 SCOPE

This procedure is applicable to calibrate high performance thin layer chromatography.

[ads id="ads1"]

3.0 RESPONSIBILITY

3.1 Doing : Technical Assistant

3.2 Checking : Executive /Manager

4.0 ACCOUNTABILITY

Head of the Department

5.0 PROCEDURE

5.1 Check the performance of instrument in terms of

1. Linearity of spotting (Annexure-I)

2. Reproducibility of spotting (Annexure-II)

5.2 Prepare the mobile phase as per method given in Annexure-I.

5.3 Prepare the solution as mentioned in Annexure-I

[ads id="ads1"]

5.4 Use 10 X 10 cm. HPTLC plate (Kieselgel 60F254 or Identical)

5.5 For Linearity of spotting

5.5.1 Apply 2 µl, 4µl, 6µl ,8µl and 10 µl of solution on HPTLC plate with spotter.

5.5.2 Allow the plate to run in mobile phase.

5.5.3 Dry the plate with drier.

5.5.4 Scan the plate with scanner.

5.5.5 Check the linearity and corelation coefficient.

5.5.6 Fill in the Data as mentioned in Annexure-I

[ads id="ads1"]

5.6 Reproducibility of spotting

5.6.1 Apply 10µl solution on HPTLC plate for five times in sequence

5.6.2 Allow the plate to run in mobile phase.

5.6.3 Dry the plate with drier.

5.6.4 Scan the plate with scanner.

5.6.5 Calculate the RSD for five track ( RSD Limit : N.M.T 3.0%)

5.6.6 Fill in the data mentioned in Annexure-II

[ads id="ads1"]

6.0 ABBREVIATIONS

6.1 %= Percentage

6.2 RSD= Relative standard deviation

6.3 N.M.T = Not more than

6.4 N.L.T = Not less than

6.5 µl = Microlitter

6.6 HPTLC = High performance thin layer chromatography

6.7 API= Active pharmaceuticals ingredient