SOP for Performance Check of Autodiluter

1.0 Objective

To check the performance of Autodiluter.

2.0 Scope

This procedure is applicable to calibrate Autodiluter.

3.0 Responsibility

3.1 Doing: Technical Assistant

3.2 Checking: Executive /Manager

[ads id="ads1"]

4.0 Accountability

Head of the Department

5.0 Procedure

5.1 Frequency: once in a 3 months

5.1 Clean the instrument with a dry cotton cloth.

5.2 Operate the instrument as per S.O.P.

5.3 For Right syringe (B)

5.3.1 Take five empty beaker, they should be conditioned & weigh previously.

5.3.2 Suck 100µl with the right syringe (B) & Note down the water weight. Further take 200µl, 300µl, 400µl & 500µl & Note down the water weight.

5.3.4 Check the linearity as shown in Annexure-I.

5.4 For left syringe (A)

5.4.1 Take five empty beaker, they should be conditioned & weigh previously.

5.4.2 Suck 1 ml with the left syringe (A) & note down the water weight.

5.4.3 Further take 2 ml, 3 ml, 4 ml, & 5 ml and note down the weight.

5.4.4 Check the linearity as shown in Annexure –II.

[ads id="ads1"]

6.0 ABBREVIATION

6.1 μl = Microliter

6.2 ml = Millilitre