SOP for Performance Check of Autodiluter
1.0 Objective
To check the performance of Autodiluter.
2.0 Scope
This procedure is applicable to calibrate Autodiluter.
3.0 Responsibility
3.1 Doing: Technical Assistant
3.2 Checking: Executive /Manager
[ads id="ads1"]
4.0 Accountability
Head of the Department
5.0 Procedure
5.1 Frequency: once in a 3 months
5.1 Clean the instrument with a dry cotton cloth.
5.2 Operate the instrument as per S.O.P.
5.3 For Right syringe (B)
5.3.1 Take five empty beaker, they should be conditioned & weigh previously.
5.3.2 Suck 100µl with the right syringe (B) & Note down the water weight. Further take 200µl, 300µl, 400µl & 500µl & Note down the water weight.
5.3.4 Check the linearity as shown in Annexure-I.
5.4 For left syringe (A)
5.4.1 Take five empty beaker, they should be conditioned & weigh previously.
5.4.2 Suck 1 ml with the left syringe (A) & note down the water weight.
5.4.3 Further take 2 ml, 3 ml, 4 ml, & 5 ml and note down the weight.
5.4.4 Check the linearity as shown in Annexure –II.
[ads id="ads1"]
6.0 ABBREVIATION
6.1 μl = Microliter
6.2 ml = Millilitre