61. What are the possible reasons for the peak shape issue of Split Peaks?

Contamination on column inlet

Incompatibility of Sample solvent with mobile phase

Elution of second component simultaneous 

Blocked frit

Column overloaded

62. What are the solutions to prevent peak shape issues of Split Peaks?

Contamination on column inlet →

1. Use guard column, or replace guard column, or replace analytical column (depending on issue identified), 

2. Backwash analytical column (less preferred method and should be rarely used)

3. If contaminants are strongly adsorbed, try to regenerate column Or 4. Replace column if issue is not resolved

Incompatibility of Sample solvent with mobile phase →

Elution of second component simultaneous → 

1. Sample cleanup before injection, 

2.Modify mobile phase composition, or 

3. Change stationary phase depending on selectivity

Blocked frit → Use in-line filter, Column backwash

Column overloaded →1. Increase column capacity by using high capability stationary

phase or increase the dimension, or, 2. Decrease injection volume

63. What are the possible reasons for the peak shape issue of Peak Fronting?

Sample solvent incompatible with mobile phase

Formation of channels in column

Low temperature of column oven 

Column overloaded

64. What are the solutions to prevent peak shape issues of Peak Fronting?

Low temperature of column oven → Increase column oven temperature to increase column temperature

Column overloaded → Use less injection volume or dilute the sample solution according to desired column  load

Formation of channels in column → Refer to the column literature to operate the system

within the given range of parameters, e.g. pH of the solution. If problem persist, replace the column

Sample diluent incompatible with mobile phase → Change the sample diluent as mobile phase

65. What are the types of Retention Time Variation or RT Variation?

Decreasing Retention Times

Increasing Retention Times

Fluctuating Retention Times

[ads id="ads1"]

66. What are the probable reasons for Decreasing Retention Times?

High flow rate

Column overloaded

Loss of bonded stationary phase

Active groups on stationary phase

Incorrect column selection

67. What are the solutions to prevent the issue of Decreasing Retention Times?

High flow rate – Check and regulate the flow rate of pump 

Column overloaded – Decrease sample injection volume, Select the column with larger internal diameter

Loss of bonded stationary phase – Replace column, Operate at recommended pH range for RP columns. Generally operated at 2-8 for silica based RP HPLC.

Active groups on stationary phase – Increase buffer strength

Incorrect column selection – Select the column with larger internal diameter

68. What are the probable reasons for Increasing Retention Times?

Changing mobile phase composition

Low flow rate 

Loss of bonding in stationary phase 

Bubbles in mobile phase 

69. What are the solutions to prevent the issue of Increasing Retention Times?

Changing mobile phase composition – 1. Cover solvent reservoirs, 2. Prepare fresh mobile phase

Low flow rate – 1. Check and adjust pump flow rate, 2. Check for leaks in system, including pump seals

Loss of bonding in stationary phase – Replace column

Bubbles in mobile phase – 1. Check flow rate and pressure, 2. Degas mobile phase

70. What are the probable reasons for Fluctuating Retention Times?

Inadequate column equilibration

Mobile phase preparation error/ composition variation

Inadequate buffer capacity

Unstable column temperature

[ads id="ads1"]

71. What are the solutions to prevent the issue of Fluctuating Retention Times?

Unstable column temperature –

1. Check column oven to ensure temperature stability, 

2. Malfunctioning of the column oven thermostat

Inadequate column equilibration – 

1. Allow columns to equilibrate for sufficient time between runs, 

2. Condition the column using concentrated sample

Mobile phase preparation error/ composition variation – 

1. Verify volume make-up of mobile phase, if needed, prepare fresh one, 

2. Verify the proportioning-valve accuracy

Inadequate buffer capacity – 

Utilize the buffer concentrations more than 20mM

72. What are the probable reasons for Ghost Peaks?

Contamination of column

Contamination of injector 

Carryover/ late eluting peak from previous injection 

Contaminated water or solvent 

Method specificity not established adequately

73. What are the solutions to prevent the issue of Ghost Peaks?

Contamination of column – 

Adequate column washing/ Flushing of column to remove contaminants

Contamination of injector –  

Injector flushing between analyses

Carryover/ late eluting peak from previous injection –

1. Increase the run time, 

2. Flush column with strong mobile phase at end of each run, 

3. For gradient runs, end the run at higher concentration 

Contaminated water or solvent – 

Use HPLC grade water or solvent

Sample contamination – 

Ensure cleanliness of glasswares, sample preparation aids such as mortar and pestle, cleanliness of sample storage area

Method specificity not established adequately –

Use sample clean-up process and

perform method specificity covering all the relevant factors (excipient peak interference, degradants, etc.)

74. What are the probable reasons for Negative Peaks?

When using RI detector, Refractive Index of solute is less than the mobile phase When using UV detector, solute absorption is less than absorption of mobile phase  Different composition of Sample solvent and mobile phase 

75. What are the solutions to prevent the issue of Negative Peaks?

When using the RI detector, the Refractive Index of solute is less than the mobile phase –

1. Use mobile phase with less refractive index, 

2. Invert detector polarity to get positive peaks

When using a UV detector, solute absorption is less than absorption of mobile phase – 

1. Change UV wavelength to get a positive peak,

 2. Identify mobile phase so that it lowers the UV absorption

Different composition of Sample solvent and mobile phase – 

Revise the composition of either sample solvent and or mobile phase to improve the compatibility and sample solubility

[ads id="ads1"]

76. What are the probable reasons for Spikes in a chromatogram?

Air bubbles in mobile phase/ inadequately degassing of mobile phase

Column stored without closing ends using column closure caps

77. What are the solutions to prevent the issue of Spikes in a chromatogram?

Air bubbles in mobile phase/ inadequately degassing of mobile phase – 

1. Degas mobile phase using suitable method, 

2. Use back pressure restrictor at detector outlet, 3. Ensure all tubings are without any leaks and fittings are tight

Column stored without closing ends using column closure caps – 

1. Ensure that column is stored with end caps closed, 

2. Flush Reverse Phase column using degassed methanol

78. What are the probable reasons for High Back Pressure in HPLC systems?

Wrong HPLC pump setting

Pressure higher during middle of gradient

Temperature lower than required

Column ageing

Column frit blockage

In-line filter blockage

Guard column blockage

System blockage

Buffer precipitation in the system

79. What are the solutions to prevent the issue of High Back Pressure in HPLC systems?

Wrong HPLC pump setting – Verify HPLC pump setting and adjust appropriately

Pressure higher during middle of gradient – This is normal phenomenon 

Temperature lower than required – Maintain the column oven temperature as per method requirement

Column ageing – It is normal phenomenon of gradual increase in pressure over the period of lifetime

Column frit blockage – 1. Column backwash, 2. Use an in-line filter to prevent column

blockage and reduce backpressure, 3. Ensure efficient filtration, 4. Use guard columns

In-line filter blockage – 1. Replace in-line filter frit with new one, 2. Centrifuge or filter samples with recommended filters, 3. Degas and Pre-filter mobile phase

Guard column blockage – Increase the replacement frequency depending on experience

Buffer precipitation in the system – Back flush the column using water

80. What are the probable reasons for Low Back Pressure in HPLC systems?

Leakage in the HPLC system

Column temperature is higher than required

Flow rate is lower than required

Next: HPLC Interview Questions -5


Previous Post Next Post

Share our Posts

Latest Learning