SOP for Calibration of Atomic Absorption Spectrophotometer (AAS)
1.0 Objective
To check the performance of the equipment for reliable and accurate results.
2.0 Scope
This procedure is applicable to check the performance of the atomic absorption spectrophotometer installed in the quality control department.
[ads id="ads1"]
3.0 Responsibility
3.1 Doing: Technical Assistant
3.2 Checking: Executive /Manager
4.0 Accountability
Head of the Department
5.0 Procedure
Frequency: Once in a month
5.1 Operate the instrument as per respective standard operating procedure.
5.2 Use Cu standard for radiations.
[ads id="ads1"]
5.3 Prepare 0.5 mcg/ml, 1.0 mcg/ml, 1.5 mcg/ml, 2.0 mcg/ml and 2.5 mcg/ml in water.
5.4 Aspirate the above solution and measure the absorbance.
5.5 Calculate the correlation coefficient by the instrument/computer.
5.6 Correlation coefficient should not less than 0.9900 %.
5.7 The performance of the instrument is satisfactory if the obtained correlation coefficient is within the given limit otherwise follow the S.O.P.
[ads id="ads1"]
6.0 ABBREVIATIONS
6.1 % = Percentage
6.2 mcg= microgram
6.3 ml= millilitre
6.4 cu= copper
[ads id="ads1"]