SOP for Calibration of pH Meter Temperature Sensor

1.0 Objective

To provide a standard operating procedure for the calibration of pH meter temperature sensor. 

2.0 Scope

This procedure is applicable to pH meter. 

3.0 Responsibility

3.1 Doing: Technical Assistant/Executive

3.2 Checking: Executive /Manager 

[ads id="ads1"]

4.0 Accountability

Head of the Department

5.0 Procedure

Frequency: Twice in Year.

5.1 Calibration Procedure for pH meter temperature sensor

5.1.1 Keep the sensor probe of reference calibrated digital thermometer in the beaker containing water for injection (WFI) / purified water (PW) of about temperature 35°C.

5.1.2 Add cold WFI/PW slowly drop wise and bring the temp. between 27°C to 30°C.

5.1.3 Compare the temp. shown by reference calibrated digital thermometer and pH meter & record it in format as Annx. I.

5.1.4 Add WFI/PW in some more quantity slowly drop wise to bring the temp. between 26°C to 23°C.

5.1.5 Compare the temp. shown by reference calibrated digital thermometer and pH meter & record it in format as Annx. I.

5.1.6 Add cold WFI/PW slowly drop wise and bring the temp. between 20°C to 22°C.

5.1.7 Compare the temp. shown by reference calibrated digital thermometer and pH meter & record it in format as Annx. I.

[ads id="ads1"]

5.1.8 Acceptance Criteria : ± 3°C