SOP for Calibration of Coulter Counter
1.0 OBJECTIVE
To calibrate the equipment for reliable and accurate results.
2.0 SCOPE
This procedure is applicable to calibrate the coulter counter installed in the quality control department.
[ads id="ads1"]
3.0 RESPONSIBILITY
3.1 Doing: Technical Assistant
3.2 Checking: Executive /Manager
4.0 ACCOUNTABILITY
Head of the Department
5.0 PROCEDURE
Frequency: Once in a three month
5.1 Program and operate the instrument as per SOP.
5.2 Use the ‘menu cursor’ keys to select the following option on the ‘SET UP’ menu.
[ads id="ads1"]
5.3 Set the parameters as follows.
Orifice size :70 & 100 µ alternately
Diameter/Length: Enter the aperture diameter and length of the orifice tube selected for calibration.
Set up: Select 'Automatic'
Analysis: Select 'Sample'
Calibration: Select 'New'
Size: Enter the (number mode) of the latex particles used for calibration and the assay Diameter of the chosen latex size should be 20 % of the aperture diameter for optimum results.(e.g. 2.15 E1 means 21.5 m)
[ads id="ads1"]
Size units: Select mm
5.4 Select ‘FULL RANGE’.
5.4.1 X-Axis ----> Select ‘Lin Dia’.
5.4.2 Y-Axis ----> Select ‘Number Diff’.
5.4.3 Blank subtract ---> Select No.
5.4.4 Channels-----> 256
5.4.5 Channel data --> No.
5.5 Observe reading as per S.O.P.
5.5.1 Move the upper and lower cursor to coincide with peak height.
5.5.2 Press SET UP key and select record on the calibration line on SET UP menu.
[ads id="ads1"]
5.5.3 The new Kd factor will display .
5.5.4 Press the CAL key, the instrument is now calibrated for that electrolyte solution and orifice tube for diameter and volume.
5.5.5 Print the graph and set up menu with new Kd factor for recording. Record the observation as per annnexure. If calibration is out of limit, follow the S.O.P.
6.0 ABBREVIATIONS
6.1 No.= Number
6.2 Lin.Dia.= Linear diameter
6.3 Diff = Difference
6.4 mm= Micrometer
6.5 m= Micron