SOP for Calibration of Plunger Operated Pipette

1.0 Objective

To calibrate the plunger operated pipette for reliable and accurate results. 

2.0 Scope

This procedure is applicable to all plunger operated pipettes used in Q. C. department.

[ads id="ads1"]

3.0 Responsibility

3.1 Doing: Technical Assistant

3.2 Checking: Executive /Manager

4.0 Accountability

Head of the Department

5.0 Procedure

Frequency: Every six month

5.1 On receipt of plunger operated pipette, perform calibration of the same.

5.2 Before testing, condition the pipette by pipetting purified water for five times and discard pipette tip.

5.3 Mount a new pipette tip and rinse with the purified water.

5.4 Weigh the beaker having a weight less than 4 gm on a microbalance.

[ads id="ads1"]

5.5 Withdraw three different volume (maximum, minimum and middle of the range) of purified water and pipette out in to dry beaker which is previously weighed.

5.6 Calibrate the pipette by wt/wt and convert to ml by using standard wt/ml at the temperature of water at the time of calibration referring Annexure –II.

5.7 Repeat the above exercise for three times and record the results in Annexure-I.

5.8 Acceptance Criteria

For 20 µl

        50 µl = + 0.7 %

        100 µl          

        200 µl

For 500 µl = + 0.5%

       1000 µl

[ads id="ads1"]

6.0 ABBREVIATIONS

6.1 SOP = Standard Operating Procedure

6.2 % = Percentage 

6.3 gm = gram 