SOP for Calibration of Automatic Box Compression Strength Tester (Digital Model)

1.0 OBJECTIVE

To calibrate the equipment for reliable and accurate results.

2.0 SCOPE

This procedure is applicable to calibrate box compression strength tester, installed in packing material testing area.

[ads id="ads1"]

3.0 RESPONSIBILITY

3.1 Doing : Technical Assistant / Executive

3.2 Checking : Executive /Manager 

4.0 ACCOUNTABILITY

Head of the department.

5.0 PROCEDURE

5.1 Frequency : Once in a month

Semi auto calibration :

[ads id="ads1"]

5.1.1 Operate the instrument as per S.O.P.

5.1.2 On zero load, press ‘ZERO’ on the digital indicator.

5.1.3 Press and hold the ‘CAL’ button, situated at the rear of the indicator. The display will show 100 kg. Now slowly place 100 kg cast iron weight in any combination.

5.1.4 Press the ‘CAL’ button again. The display will show CAL – F. This denotes completion of calibration. Press the ‘CAL’ button once again. The display should show 100 ± 1kg.

5.1.5 Note down the observation in Annexure – I.

[ads id="ads1"]

5.1.6 Remove the cast iron weights carefully.

5.2 Manual check calibration.

5.2.1 After completion of semiauto calibration take out 20 kg each cast iron weight from 100 kg & note the weight for 80 kg calibration. Similarly take out gradually each 20 kg of weight one after other & note the weight for 60, 40 & 20 kgs calibration.

5.2.2 Note down the observation in Annexure – I.

5.3 If instrument found out of calibration follow the SOP for Out of Calibration.

[ads id="ads1"]

6.0 ABBREVIATIONS

6.1 SOP - Standard Operating Procedure

6.1 kg - kilogram


[ads id="ads1"]

Previous Post Next Post

Share our Posts

Latest Learning